4.2 Setting Basic Settings of Measurement Conditions
42
Setting the measurement frequency may change the setting for the measurement signal
level.
See "For setting range and accuracy" (p. 45)
4
Use the numeric keypad to enter the
frequency.
5
Press a unit key to confirm the setting.
6
Press to close the setting screen.
Use or to select the posi-
tion of the decimal point and the unit.
The frequency is not confirmed until a unit key is
pressed.
The unit keys are disabled until a number is
entered.
If you attempt to set a test frequency greater
than 5 MHz, it will automatically be reduced to 5
MHz.
If you attempt to set a test frequency lower than
4 Hz, it will automatically be increased to 4 Hz.
See (p. 387)
Sets the measurement frequency to x10.
Sets the measurement frequency to x1/10.
If you make a mistake during input:
Press the to cancel the input so far, and
start again.
Frequency Setting Frequency Setting
Frequency Setting Frequency Setting
Use or to enter each digit of the
frequency.
Holding down a digit key changes the value continuously.