REV 0, June 2005 Page 327
Operation Manual SUPREMA
12. Sensor Data Sheets
The left-hand cursor key has to be activated in the input field limit value (2
nd
to 4
th
alarm) as long as only zeros are displayed. Then press press the lower cursor key:
the field content is deleted (indication: ”——“), the alarm is deactivated (PC
operation software: to delete field inputs (DEL key, then confirm by the ENTER
key).
All other inputs in the Settings/Measuring Points menu may be chosen freely.
Note: The potential-free contacts can also be used to carry out a lamp test or
relay status of contacts. Use of these contacts for this purpose is identical to an
alarm and will also trigger the collective alarm.