5.7 Judging Measurement Results (Comparator Function)
202
Set the reference value, upper limit value, and lower limit
value.
If you make a mistake during input:
press to cancel the input and start again.
1. Press and use the numeric keypad to set the
reference value.
Settable range: -9.999999G to 9.999999G
2. Press to confirm the reference value.
3. Press and use the numeric keypad to set the
upper limit value.
Settable range: -999.9999 to 999.9999
(Set as a % value)
Settable range: -9.999999G to 9.999999G
(Set as a reference value)
4. Press to confirm the upper limit value.
5. Press and use the numeric keypad to set the
lower limit value.
Settable range: -999.9999 to 999.9999
(Set as a % value)
Settable range: -9.999999G to 9.999999G
(Set as a reference value)
6. Press to confirm the lower limit value.
7. Press to confirm the setting.
Upper/Lower Limit & Reference Value Settings