EasyManua.ls Logo

Hioki IM3570 - Page 54

Hioki IM3570
458 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.2 Setting Basic Settings of Measurement Conditions
46
When Low Z high accuracy mode (p. 54)
Measurement frequency
setting range
Open circuit voltage
setting range
Open circuit voltage
accuracy
4 Hz to 100 kHz 0.005 V to 1 V ±10%±10 mV
4 Hz 5 MHz100 kHz
0.005 V
1.000 V
Testing some types of sample is not possible using constant voltage. In this case, the following
symbol appears on the display:
In such a case, constant voltage testing is not performed. Change the constant voltage level
to a value not more than the value being shown as Vmoni.
When a 1 F capacitance is measured at 10 kHz, the CV operation range can be obtained
as follows. Sample impedance
Zm becomes as follows:
Sample impedance
Zm becomes as follows:
The impedance
Zm' observed from the generator is as follows:
Accordingly, the voltage
Vm across both leads of the sample is as follows:
Because the generator output voltage range is 5[mV] to 5[V] for 10 kHz, the CV operation
range per the above expression is
Vm = 0.8[mV] to 0.78[V].
In low Z high accuracy mode, the output resistance
Ro becomes 10[].
The open circuit voltage that can be set differs depending on the frequency.
When the test frequency is higher than 1 MHz at an open-circuit voltage of more than 1 V, the
voltage is automatically set to 1 V.
When the measurement frequency is set higher than 1 MHz at an open circuit voltage of less than 0.010 V,
the voltage is automatically set to 0.010 V.
Xm
1
2
fC
-----------------=
Zm = Rm + jXm = 0 [

] - j15.9 [

]
Ro: Output resistance (100 [])Zm' = Ro + Zm = 100 [

] - j15.9 [

]
Vm
Zm Vo
Zm
----------------------
15.9
Vo
101.3

--------------------------
==
Vo : generator output
|Zm' |
15.9 [

]×Vo

Table of Contents

Other manuals for Hioki IM3570

Related product manuals