1136 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:USB:TRIGger:SEU
(see page 1610)
Command Syntax
:SBUS<n>:USB:TRIGger:SEU <string>
<string> ::= "nn" where n ::= {0 | 1 | X}
<string> ::= "0xn" where n ::= {0 | 1 |2|3|X|$}
The :SBUS<n>:USB:TRIGger:SEU command specifies the 2-bit S and E or U
portion of the trigger for the selected bus, in binary or hex.
Query Syntax
:SBUS<n>:USB:TRIGger:SEU?
The :SBUS<n>:USB:TRIGger:SEU? query returns the specified S and E or U portion
of the trigger.
Return Format
<string><NL>
<string> ::= "nn" where n ::= {0 | 1 | X}
See Also • ":SBUS<n>:USB:TRIGger:PID:SPECial" on page 1132