:SBUS<n> Commands 29
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 763
:SBUS<n>:A429:BASE
(see page 1428)
Command Syntax
:SBUS<n>:A429:BASE <base>
<base> ::= {BINary | HEX}
The :SBUS<n>:A429:BASE command selects between hexadecimal and binary
display of the decoded data.
The BASE command has no effect on the SDI and SSM fields, which are always
displayed in binary, nor the Label field, which is always displayed in octal.
Query Syntax
:SBUS<n>:A429:BASE?
The :SBUS<n>:A429:BASE? query returns the current ARINC 429 base setting.
Return Format
<base><NL>
<base> ::= {BIN | HEX}
Errors "-241, Hardware missing" on page 1387
See Also "Introduction to :SBUS<n> Commands" on page 755
":SBUS<n>:MODE" on page 759
":SBUS<n>:A429:FORMat" on page 767