1154 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
33 :TRIGger Commands
:TRIGger:DELay:TDELay:TIME
(see page 1428)
Command Syntax
:TRIGger:DELay:TDELay:TIME <time_value>
<time_value> ::= time in seconds in NR3 format
The :TRIGger:DELay:TDELay:TIME command sets the delay time between the
arming edge and the trigger edge in the Edge Then Edge trigger. The time is in
seconds and must be from 4 ns to 10 s.
Query Syntax
:TRIGger:DELay:TDELay:TIME?
The :TRIGger:DELay:TDELay:TIME? query returns current delay time setting.
Return Format
<time value><NL>
<time_value> ::= time in seconds in NR3 format
See Also "Introduction to :TRIGger Commands" on page 1139
":TRIGger:DELay:TRIGger:SLOPe" on page 1156