966 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:UART:TRIGger:TYPE
(see page 1428)
Command Syntax
:SBUS<n>:UART:TRIGger:TYPE <value>
<value> ::= {RSTArt | RSTOp | RDATa | RD1 | RD0 | RDX | PARityerror
| TSTArt | TSTOp | TDATa | TD1 | TD0 | TDX}
The :SBUS<n>:UART:TRIGger:TYPE command selects the UART trigger type.
When one of the RD or TD types is selected, the :SBUS<n>:UART:TRIGger:DATA
and :SBUS<n>:UART:TRIGger:QUALifier commands are used to specify the data
value and comparison operator.
The RD1, RD0, RDX, TD1, TD0, and TDX types (for triggering on data and alert bit
values) are only valid when a 9-bit width has been selected.
Query Syntax
:SBUS<n>:UART:TRIGger:TYPE?
The :SBUS<n>:UART:TRIGger:TYPE? query returns the current UART trigger data
value.
Return Format
<value><NL>
<value> ::= {RSTA | RSTO | RDAT | RD1 | RD0 | RDX | PAR | TSTA |
TSTO | TDAT | TD1 | TD0 | TDX}
See Also "Introduction to :TRIGger Commands" on page 1139
":TRIGger:MODE" on page 1148
":SBUS<n>:UART:TRIGger:DATA" on page 963
":SBUS<n>:UART:TRIGger:QUALifier" on page 965
":SBUS<n>:UART:WIDTh" on page 967