:WGEN<w> Commands 35
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1295
:WGEN<w>:MODulation:FUNCtion
(see page 1428)
Command Syntax
:WGEN<w>:MODulation:FUNCtion <shape>
<w> ::= 1 in NR1 format
<shape> ::= {SINusoid | SQUare| RAMP}
The :WGEN<w>:MODulation:FUNCtion command specifies the shape of the
modulating signal.
When the RAMP shape is selected, you can specify the amount of time per cycle
that the ramp waveform is rising with the
:WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry command.
This command applies to AM and FM modulation. (The FSK modulation signal is a
square wave shape.)
Query Syntax
:WGEN<w>:MODulation:FUNCtion?
The :WGEN<w>:MODulation:FUNCtion? query returns the specified modulating
signal shape.
Return Format
<shape><NL>
<shape> ::= {SIN | SQU| RAMP}
See Also • ":WGEN<w>:MODulation:AM:DEPTh" on page 1289
• ":WGEN<w>:MODulation:AM:FREQuency" on page 1290
• ":WGEN<w>:MODulation:FM:DEViation" on page 1291
• ":WGEN<w>:MODulation:FM:FREQuency" on page 1292
• ":WGEN<w>:MODulation:FSKey:FREQuency" on page 1293
• ":WGEN<w>:MODulation:FSKey:RATE" on page 1294
• ":WGEN<w>:MODulation:FUNCtion:RAMP:SYMMetry" on page 1296
• ":WGEN<w>:MODulation:STATe" on page 1298
• ":WGEN<w>:MODulation:TYPE" on page 1299