232 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
6 Root (:) Commands
:AUToscale:CHANnels
(see page 1428)
Command Syntax
:AUToscale:CHANnels <value>
<value> ::= {ALL | DISPlayed}
The :AUToscale:CHANnels command specifies which channels will be displayed on
subsequent :AUToscales.
When ALL is selected, all channels that meet the requirements of :AUToscale
will be displayed.
When DISPlayed is selected, only the channels that are turned on are
autoscaled.
Use the :VIEW or :BLANk root commands to turn channels on or off.
Query Syntax
:AUToscale:CHANnels?
The :AUToscale:CHANnels? query returns the autoscale channels setting.
Return Format
<value><NL>
<value> ::= {ALL | DISP}
See Also "Introduction to Root (:) Commands" on page 226
":AUToscale" on page 229
":AUToscale:AMODE" on page 231
":VIEW" on page 262
":BLANk" on page 234