EasyManua.ls Logo

Keysight Technologies InfiniiVision MSO-X 4022A

Keysight Technologies InfiniiVision MSO-X 4022A
1580 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Status Reporting 39
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1401
Trigger Event Register (TER)
This register sets the TRG bit in the status byte when a trigger event occurs.
The TER event register stays set until it is cleared by reading the register or using
the *CLS command. If your application needs to detect multiple triggers, the TER
event register must be cleared after each one.
If you are using the Service Request to interrupt a program or controller operation,
you must clear the event register each time the trigger bit is set.

Table of Contents

Related product manuals