1226 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
33 :TRIGger Commands
:TRIGger:ZONE:SOURce
(see page 1428)
Command Syntax
:TRIGger:ZONE:SOURce <source>
<source> ::= {CHANnel<n>}
<n> ::= 1 to (# analog channels) in NR1 format
The :TRIGger:ZONE:SOURce command sets the analog source channel shared by
all zones.
Query Syntax
:TRIGger:ZONE:SOURce?
The :TRIGger:ZONE:SOURce? query returns the analog source channel specified
for zone qualified triggers.
Return Format
<source><NL>
<source> ::= {CHAN<n>}
See Also ":TRIGger:ZONE:STATe" on page 1227