1228 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
33 :TRIGger Commands
:TRIGger:ZONE<n>:MODE
(see page 1428)
Command Syntax
:TRIGger:ZONE<n>:MODE <mode>
<mode> ::= {INTersect | NOTintersect}
<n> ::= 1-2 in NR1 format
The :TRIGger:ZONE<n>:MODE command sets the zone qualifying condition for
Zone 1 or Zone 2 as either "Must Intersect" or "Must Not Intersect".
Query Syntax
:TRIGger:ZONE<n>:MODE?
The :TRIGger:ZONE<n>:MODE? query returns the zone qualifying condition for
Zone 1 or Zone 2.
Return Format
<mode><NL>
<mode> ::= {INT | NOT}
See Also ":TRIGger:ZONE<n>:STATe" on page 1231
":TRIGger:ZONE<n>:PLACement" on page 1229
":TRIGger:ZONE<n>:VALidity" on page 1230