604 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
24 :MTESt Commands
:MTESt:RMODe:TIME
(see page 1428)
Command Syntax
:MTESt:RMODe:TIME <seconds>
<seconds> ::= from 1 to 86400 in NR3 format
When the :MTESt:RMODe command is set to TIME, the :MTESt:RMODe:TIME
command sets the number of seconds for a mask test to run.
Query Syntax
:MTESt:RMODe:TIME?
The :MTESt:RMODe:TIME? query returns the number of seconds currently set.
Return Format
<seconds><NL>
<seconds> ::= from 1 to 86400 in NR3 format
See Also "Introduction to :MTESt Commands" on page 581