1254 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
34 :WAVeform Commands
:WAVeform:SEGMented:TTAG
(see page 1428)
Query Syntax
:WAVeform:SEGMented:TTAG?
The :WAVeform:SEGMented:TTAG? query returns the time tag of the currently
selected segmented memory index. The index is selected using the
:ACQuire:SEGMented:INDex command.
Return Format
<time_tag> ::= in NR3 format
See Also ":ACQuire:SEGMented:INDex" on page 272
"Introduction to :WAVeform Commands" on page 1235
Example Code "Example Code" on page 272
NOTE
This command is available when the segmented memory option (Option SGM) is enabled.