:MARKer Commands 21
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 457
:MARKer:X1Position
(see page 1428)
Command Syntax
:MARKer:X1Position <position> [suffix]
<position> ::= X1 cursor position in NR3 format
<suffix> ::= {s | ms | us | ns | ps | Hz | kHz | MHz}
The :MARKer:X1Position command:
Sets :MARKer:MODE to MANual if it is not currently set to WAVeform (see
":MARKer:MODE" on page 455).
Sets the X1 cursor position to the specified value.
X cursor units are set by the :MARKer:XUNits command.
Query Syntax
:MARKer:X1Position?
The :MARKer:X1Position? query returns the current X1 cursor position. This is
functionally equivalent to the obsolete :MEASure:TSTArt command/query.
Return Format
<position><NL>
<position> ::= X1 cursor position in NR3 format
See Also "Introduction to :MARKer Commands" on page 453
":MARKer:MODE" on page 455
":MARKer:X2Position" on page 460
":MARKer:X1Y1source" on page 458
":MARKer:X2Y2source" on page 461
":MARKer:XUNits" on page 463
":MEASure:TSTArt" on page 1360
NOTE
If the front-panel cursors are off, the marker position values are not defined and an error is
generated. Make sure to set :MARKer:MODE to MANual or WAVeform to put the cursors in the
front-panel Normal mode.