:MEASure Commands 22
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 519
:MEASure:PHASe
(see page 1428)
Command Syntax
:MEASure:PHASe [<source1>][,<source2>]
<source1>, <source2> ::= {CHANnel<n> | FUNCtion<m> | MATH<m> | WMEMory<r
>}
<n> ::= 1 to (# analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:PHASe command places the instrument in the continuous
measurement mode and starts a phase measurement.
Query Syntax
:MEASure:PHASe? [<source1>][,<source2>]
The :MEASure:PHASe? query measures and returns the phase between the
specified sources.
A phase measurement is a combination of the period and delay measurements.
First, the period is measured on source1. Then the delay is measured between
source1 and source2. The edges used for delay are the source1 rising edge used
for the period measurement closest to the horizontal reference and the rising edge
on source 2. See :MEASure:DELay for more detail on selecting the 2nd edge.
The phase is calculated as follows:
phase = (delay / period of input 1) x 360
Return Format
<value><NL>
<value> ::= the phase angle value in degrees in NR3 format
See Also "Introduction to :MEASure Commands" on page 488
":MEASure:DELay" on page 500
":MEASure:PERiod" on page 518
":MEASure:SOURce" on page 529