Obsolete and Discontinued Commands 37
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1375
:PRINt?
(see page 1428)
Query Syntax
:PRINt? [<options>]
<options> ::= [<print option>][,..,<print option>]
<print option> ::= {COLor | GRAYscale | BMP8bit | BMP}
The :PRINt? query pulls image data back over the bus for storage.
NOTE
The :PRINT command is an obsolete command provided for compatibility to previous
oscilloscopes. Use the :DISPlay:DATA command (see page 369) instead.
Print Option :PRINt command :PRINt? query Query Default
COLor Sets palette=COLor
GRAYscale Sets
palette=GRAYscale
palette=COLor
PRINter0,1 Causes the USB
printer #0,1 to be
selected as
destination (if
connected)
Not used N/A
BMP8bit Sets print format to
8-bit BMP
Selects 8-bit BMP
formatting for query
N/A
BMP Sets print format to
BMP
Selects BMP
formatting for query
N/A
FACTors Selects outputting of
additional settings
information for :PRINT
Not used N/A
NOFactors Deselects outputting
of additional settings
information for :PRINT
Not used N/A
Old Print Option: Is Now:
HIRes COLor
LORes GRAYscale
PARallel PRINter0
DISK invalid
PCL invalid