678 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
26 :POWer Commands
:POWer:PSRR:WGEN:VOLTage:PROFile
(see page 1428)
Command Syntax
:POWer:PSRR:WGEN:VOLTage:PROFile {{0 | OFF} | {1 | ON}}
The :POWer:PSRR:WGEN:VOLTage:PROFile command enables or disables the
ability to set initial waveform generator ramp amplitudes for each frequency range.
With amplitude profiling, you can use lower amplitudes at frequencies where the
device under test (DUT) is sensitive to distortion and use higher amplitudes where
the DUT is less sensitive to distortion.
You can often observe distortions during the test. If the input test sine wave begins
to look lopsided, clipped, or somewhat triangular in shape (nonsinusoidal), you are
probably encountering distortion due to overdriving your DUT. Optimizing test
amplitudes to achieve the best dynamic range measurements is often an iterative
process of running your frequency response measurements multiple times.
Query Syntax
:POWer:CLResponse:WGEN:VOLTage:PROFile?
The :POWer:PSRR:WGEN:VOLTage:PROFile? query returns the voltage profile
setting.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also ":POWer:PSRR" on page 667
":POWer:PSRR:APPLy" on page 668
":POWer:PSRR:DATA" on page 669
":POWer:PSRR:FREQuency:MAXimum" on page 670
":POWer:PSRR:FREQuency:MINimum" on page 671
":POWer:PSRR:FREQuency:MODE" on page 672
":POWer:PSRR:PPDecade" on page 673
":POWer:PSRR:SOURce:INPut" on page 674
":POWer:PSRR:SOURce:OUTPut" on page 675
":POWer:PSRR:WGEN:LOAD" on page 676
":POWer:PSRR:WGEN:VOLTage" on page 677