EasyManua.ls Logo

Keysight Technologies InfiniiVision MSO-X 4022A

Keysight Technologies InfiniiVision MSO-X 4022A
1580 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
562 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
23 :MEASure Power Commands
:MEASure:CRESt
(see page 1428)
Command Syntax
:MEASure:CRESt [<source>]
<source> ::= {CHANnel<n>| FUNCtion<m> | MATH<m>}
<n> ::= 1 to (# analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
The :MEASure:CRESt command installs a crest factor measurement on screen.
The <source> parameter is the channel probing current or voltage. This source can
also be specified by the :MEASure:SOURce command.
Crest factor is a measure of power quality. It is the ratio between the
instantaneous peak AC line current (or voltage) required by the load and the RMS
current (or voltage). For example: Ipeak / IRMS or Vpeak / VRMS.
Query Syntax
:MEASure:CRESt? [<source>]
The :MEASure:CRESt query returns the measured crest factor.
Return Format
<return_value><NL>
<return_value> ::= the crest factor value in NR3 format
See Also ":MEASure:SOURce" on page 529
":POWer:QUALity:APPLy" on page 679

Table of Contents

Related product manuals