:SBUS<n> Commands 29
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 861
:SBUS<n>:I2S:WSLow
(see page 1428)
Command Syntax
:SBUS<n>:I2S:WSLow <low_def>
<low_def> ::= {LEFT | RIGHt}
The :SBUS<n>:I2S:WSLow command selects the polarity of the word select (WS)
signal:
LEFT— a word select (WS) state of low indicates left channel data is active on
the I2S bus, and a WS state of high indicates right channel data is active on the
bus.
RIGHt — a word select (WS) state of low indicates right channel data is active
on the I2S bus, and a WS state of high indicates left channel data is active on
the bus.
Query Syntax
:SBUS<n>:I2S:WSLow?
The :SBUS<n>:I2S:WSLow? query returns the currently selected I2S word select
(WS) polarity.
Return Format
<low_def><NL>
<low_def> ::= {LEFT | RIGHt}
See Also "Introduction to :TRIGger Commands" on page 1139
":SBUS<n>:I2S:ALIGnment" on page 845
":SBUS<n>:I2S:CLOCk:SLOPe" on page 847
":SBUS<n>:I2S:RWIDth" on page 848
":SBUS<n>:I2S:TWIDth" on page 860