:TRIGger Commands 33
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1177
:TRIGger:GLITch:RANGe
(see page 1428)
Command Syntax
:TRIGger:GLITch:RANGe <less_than_time>[suffix],
<greater_than_time>[suffix]
<less_than_time> ::= (15 ns - 10 seconds) in NR3 format
<greater_than_time> ::= (10 ns - 9.99 seconds) in NR3 format
[suffix] ::= {s | ms | us | ns | ps}
The :TRIGger:GLITch:RANGe command sets the pulse width duration for the
selected :TRIGger:GLITch:SOURce. You can enter the parameters in any order —
the smaller value becomes the <greater_than_time> and the larger value becomes
the <less_than_time>.
Query Syntax
:TRIGger:GLITch:RANGe?
The :TRIGger:GLITch:RANGe? query returns the pulse width duration time for
:TRIGger:GLITch:SOURce.
Return Format
<less_than_time>,<greater_than_time><NL>
See Also "Introduction to :TRIGger Commands" on page 1139
":TRIGger:GLITch:SOURce" on page 1178
":TRIGger:GLITch:QUALifier" on page 1176
":TRIGger:MODE" on page 1148