964 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:UART:TRIGger:IDLE
(see page 1428)
Command Syntax
:SBUS<n>:UART:TRIGger:IDLE <time_value>
<time_value> ::= time from 1 us to 10 s in NR3 format
The :SBUS<n>:UART:TRIGger:IDLE command selects the value of the idle period
for burst trigger in the range from 1 us to 10 s when in UART mode.
Query Syntax
:SBUS<n>:UART:TRIGger:IDLE?
The :SBUS<n>:UART:TRIGger:IDLE? query returns the current UART trigger idle
period time.
Return Format
<time_value><NL>
<time_value> ::= time from 1 us to 10 s in NR3 format
See Also "Introduction to :TRIGger Commands" on page 1139
":TRIGger:MODE" on page 1148
":SBUS<n>:UART:TRIGger:BURSt" on page 962
":SBUS<n>:UART:TRIGger:TYPE" on page 966