902 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
29 :SBUS<n> Commands
:SBUS<n>:SENT:DISPlay
(see page 1428)
Command Syntax
:SBUS<n>:SENT:DISPlay <base>
<base> ::= {HEX | DECimal | SYMBolic}
The :SBUS<n>:SENT:DISPlay command specifies the number base used by the
decoder. The chosen base is used for the data nibbles in Raw decode format, the
defined Signals in the other formats, and for the data field of the Serial Messages.
This selection is used for both the Lister and the decode line displays.
When SYMBolic is selected, Fast Channel Signals display a calculated physical
value based on the specified multiplier and offset:
PhysicalValue = (Multiplier * SignalValueAsUnsignedInteger) + Offset
When SYMBolic is selected, the CRC and Slow Channel information is displayed in
hex.
Query Syntax
:SBUS<n>:SENT:DISPlay?
The :SBUS<n>:SENT:DISPlay? query returns the SENT decode number base
setting.
Return Format
<base><NL>
<base> ::= {HEX | DEC | SYMB}
See Also ":SBUS<n>:SENT:CLOCk" on page 900
":SBUS<n>:SENT:CRC" on page 901
":SBUS<n>:SENT:FORMat" on page 904
":SBUS<n>:SENT:IDLE" on page 906
":SBUS<n>:SENT:LENGth" on page 907
":SBUS<n>:SENT:PPULse" on page 908
":SBUS<n>:SENT:SIGNal<s>:DISPlay" on page 909
":SBUS<n>:SENT:SIGNal<s>:LENGth" on page 910
":SBUS<n>:SENT:SIGNal<s>:MULTiplier" on page 912
":SBUS<n>:SENT:SIGNal<s>:OFFSet" on page 913
":SBUS<n>:SENT:SIGNal<s>:ORDer" on page 914
":SBUS<n>:SENT:SIGNal<s>:STARt" on page 916
":SBUS<n>:SENT:SOURce" on page 918
":SBUS<n>:SENT:TOLerance" on page 920
":SBUS<n>:SENT:TRIGger" on page 921
":SBUS<n>:SENT:TRIGger:FAST:DATA" on page 923