EasyManua.ls Logo

Keysight Technologies InfiniiVision MSO-X 4022A

Keysight Technologies InfiniiVision MSO-X 4022A
1580 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Synchronizing Acquisitions 40
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1421
myScope.WriteString ":OPERegister:CONDition?"
varQueryResult = myScope.ReadNumber
' Mask RUN bit (bit 3, &H8).
If (varQueryResult And &H8) = 0 Then
Exit Do
Else
Sleep 100 ' Small wait to prevent excessive queries.
lngElapsed = lngElapsed + 100
End If
Loop
' Get results.
' -----------------------------------------------------------------
If lngElapsed < lngTimeout Then
myScope.WriteString ":MEASure:RISetime"
myScope.WriteString ":MEASure:RISetime?"
varQueryResult = myScope.ReadNumber ' Read risetime.
Debug.Print "Risetime: " + _
FormatNumber(varQueryResult * 1000000000, 1) + " ns"
Else
Debug.Print "Timeout waiting for single-shot trigger."
End If
Exit Sub
VisaComError:
MsgBox "VISA COM Error:" + vbCrLf + Err.Description
End Sub

Table of Contents