:MEASure Commands 22
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 501
90%, 50%, and 10% values between Vbase and Vtop. If you want to move the
delay measurement point nearer to Vtop or Vbase, you must change the threshold
values with the :MEASure:DEFine THResholds command.
Return Format
<value><NL>
<value> ::= floating-point number delay time in seconds in NR3 format
See Also "Introduction to :MEASure Commands" on page 488
":MEASure:DEFine" on page 497
":MEASure:PHASe" on page 519