MEASurement Commands 2
N1911A/1912A P-Series Power Meters Programming Guide 121
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command sub-systems. The parameters must
be entered in the specified order. If parameters are omitted, they default from the right. The parameter DEFault is used as a place holder.
Specifying DEF leaves the parameter value unchanged.
[b] When the measurement result is linear this parameter represents the number of significant digits. When the measurement result is
logarithmic 1 to 4 represents of 1, 0.1, 0.01 and 0.001 respectively.
[c] N1912A only.
Example
source list The channel which the command is implemented on.
If unspecified the current window setup is used. However,
on the N1912A, if the window shows a ratio or difference
measurement, the upper window defaults to Channel A and
the lower window to Channel B.
(@1)
(@2)
[c]
CONF1:POW:AC DEF,2,(@1) This command configures the upper
window/upper measurement to measure
the power of Channel A, using the current
sensor range and a resolution setting of 2.