10 SENSe Subsystem
442 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:TIME <numeric_value>
This command sets the length of the time-gated period (time-gate length) for
time-gated measurements for the P-Series and E9320 sensors which are set to
NORMal mode. To set an E9320 sensor to NORMal mode, refer to the command
“[SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>” on page 414.
Syntax
Parameters
Item Description/Default Range of Values
numeric_value The length of the time gated period in seconds.
DEF: the default value is 100 µs
Units are resolved to 1 ns.
0 to 1 second
DEF