OUTPut Subsystem 8
N1911A/1912A P-Series Power Meters Programming Guide 309
OUTPut:RECorder[1]|2:LIMit:UPPer <numeric_value>
This command sets the maximum scaling value for the specified recorder output.
The units used are dependent on the units currently set for the CALC block
specified in OUTPut:RECorder[1]|2:FEED <data_handle>.
Syntax
Parameters
Example
Reset Condition
On reset, the maximum scaling value is set to +20 dBm.
Query
OUTPut:RECorder[1]|2:LIMit:UPPer?
The query command returns the maximum scaling value.
Item Description/Default Range of Values
numeric_value A numeric value for the minimum scaling value. The units
used—dBm, W or %—are dependent on the units currently
set for the CALC block specified in
OUTPut:RECorder[1]|2:FEED <data_handle>.
–150 to +230 dBm
1 aW to 100 XW
0 % to 999 %
OUTP:REC:LIM:UPP 10 This command sets the maximum scaling
value to 10.
OUTP
:LIM
:REC
Space
?
:UPP
numeric_value
1
2