10 SENSe Subsystem
430 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:POWer:AC:RANGe <numeric_value>
This command is only valid when used with an E-Series power sensor. Its purpose
is to select one of two power ranges.
If 0 is selected, the power sensor’s lower range is selected
If 1 is selected, the power sensor’s upper range is selected
Setting a range with this command automatically switches
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO to OFF.
Syntax
Example
Reset Condition
On reset, the upper range is selected.
Query
[SENSe[1]]|SENSe2:POWer:AC:RANGe?
The query enters a 1 or 0 into the output buffer indicating the status of the power
sensor’s range.
1 is returned when the upper range is selected
0 is returned when the lower range is selected
POW:AC:RANG 0 This command sets the power sensor to it’s
lower range.
SENS
1
:AC
POW
SENS2
:
:RANG
Space
?
numeric_value