17 IEEE 488.2 Command Reference
744 N1911A/1912A P-Series Power Meters Programming Guide
*RCL <NRf>
The *RCL <NRf> (ReCaLl) command restores the state of the power meter from
the specified save/recall register. An instrument setup must have been stored
previously in the specified register.
Syntax
Parameters
Error Message
If the register does not contain a saved state, error –221, “Settings conflict”
occurs.
Type Description/Defaul t Range of Values
NRf The number of the register to be recalled. 1 - 10
*RCL
NRf
Space