10 SENSe Subsystem
424 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed]:STOP
<numeric_value>
This command sets the stop frequency of average or peak frequency sweep. It
must be used in conjunction with external trigger.
If frequency sweep is disabled (frequency sweep step set to 0), stop frequency will
be set but will not take effect.
NOTE
This command is only applicable when used with E4410, N8480 (exclduing
Option CFT), E9300, E9320 or N1920 sensor.
NOTE
SENS:FREQ:STAR, SENS:FREQ:STOP and SENS:FREQ:STEP are allowed to be set in
any desirable sequence.
When frequency sweep mode is configured with frequency step size within its
allowable range, 1 to 2048, the following applies:
If frequency stop point is greater than frequency start point, the frequency
range will be sweep in an ascending order.
If frequency stop point is less than frequency start point, the frequency range
will be sweep in a descending order.
If frequency stop point and frequency start point are equal, it is the same as
power sweep mode.