5 DISPlay Subsystem
246 N1911A/1912A P-Series Power Meters Programming Guide
DISPlay[:WINDow[1]|2]:METer:LOWer <numeric_value>
This command sets the analog meter lower scale limit.
The units used are dependent on the current setting of UNIT:POWer and
CALCulate:RELative:STATe as shown in Table 5-3.
Table 5-3 Measurement Units
NOTE
This command has the same purpose as DISPlay[:WINDow[1]|2]:ANALog:LOWer
<numeric_value>.
Measurement Mode Measurement Type CALC:REL:STAT OFF CALC:REL:STAT ON
Linear Log Linear Log
Single Channel
Avg, Pk Watt dBm % dB
Pk–Avg % dB % dB
Ratio
Avg, Pk, Pk–Avg % dB % dB
Difference
Avg, Pk Watt dBm % dB
Pk–Avg % dB % dB

Table of Contents