PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 361
PSTatistic[1]|2:CCDF:TRACe:POWer:PEAK?
This command is used to retrieve peak power value of Channel A or Channel B
trace.
Syntax
Example
Error Messages
If no power sensor is connected, error –241 "Hardware missing" occurs.
If sensor/sensors connected are not P-Series sensors, error –241 "Hardware
missing" occurs.
If measurement speed setting is FAST, error –221 "Settings conflict" occurs.
If command executed in other window besides CCDF window, error –221
"Settings conflict: Requires CCDF window" occurs.
If the acquisition mode is in continuous triggering or triggering source is set to
either INT1, INT2, or EXT, error –221 "Settings conflict" occurs.
NOTE
This command is only applicable when P-Series sensors are present and the
following conditions are met:
Free-run acquisition mode is selected
NORMal or DOUBle measurement speed setting is chosen
:POW
PST
2
1
:CCDF :TRAC
?
:PEAK
PST:CCDF:TRAC:POW:PEAK? This command returns the peak power value
of Channel A trace.