SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 381
[SENSe[1]]|SENSe2:AVERage2:COUNt <numeric_value>
This command is used to enter the video filter length for the P-Series and E9320
sensor. Video filtering is applied to the traces. Successive traces are combined to
reduce noise without affecting the dynamic characteristic of the signal.
Syntax
Parameters
[a] This is only implemented in powers of 2 (2
n
).
Example
Item Description/Default Range of Values
numeric_value A numeric value defining the filter length.
DEF: the default value is 4.
1 to 256
[a]
DEF
AVER2:COUN 16 This command enters a video filter length of
16 for Channel A.
:COUN
Space
?
numeric_value
DEF
SENS
1
AVER2
SENS2
: