10 SENSe Subsystem
448 N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:V2P ATYPe|DTYPe
This command is used to select the type of linearity correction that is applied to
the channel sensors being used. For most 8480 Series sensors, the correct (A type
or D type) linearity correction table is automatically selected. However, for the
V8486A and W8486A sensors, D type (diode) correction is selected and the
automatic selection must be overridden.
Syntax
Example
Reset Condition
On reset, the linearity correction is set for A type.
Query
[SENSe[1]]|SENSe2:V2P?
The query returns the current type of linearity correction being displayed on the
screen.
This command is only applicable for V8486A and W8486A sensors.
SENS2:V2P DTYP This command selects the D type linearity
correction to be applied to Channel B.