SERVice Subsystem 15
N1911A/1912A P-Series Power Meters Programming Guide 705
SERVice:SENSor[1]|2:CALFactor <cal_factor_data>
This command writes calibration factor data to, or reads calibration factor data
from, the currently connected sensor. The whole calibration factor block must be
written at once as a checksum is generated. The new block must not be larger
than the existing block.
This command applies to the following sensors:
E4410 Series
N8480 Series (excluding Option CFT)
E9300 Series
E9320 Series, average path data
For E9320 Series sensors, peak path, refer to “SERVice:SENSor[1]|2:PCALfactor
<cal_factor_data>” on page 713.
Syntax
Parameters
Item Description/Default
cal_factor_data A binary data block. Refer to Appendix A, “Calibration Factor Block Layout,” on page
759 for further information.