SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 439
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:MARKer[1]|2:POWer?
This query is used to retrieve the power measurement at marker 1 or marker 2
specified for the gate.
The following settings are required to use the command:
–Sensor mode: Peak
Trigger mode: Single INT/EXT trigger (INIT:CONT OFF) or continuous
INT/EXT trigger (INIT:CONT ON)
Syntax
Example
Error Messages
If the query is used when the following conditions are not met, error –221,
“Settings conflict” occurs:
–Sensor mode: Peak
Trigger mode: Single INT/EXT trigger (INIT:CONT OFF) or continuous
INT/EXT trigger (INIT:CONT ON)
SENS:SWE2:MARK1:POW? This query returns the power measurement
at marker 1 of Gate 2 of Channel A.
SENS
1
SENS2
:
:MARK
SWE
?
1
2
3
4
1
2
:POW