1 Power Meter Remote Operation
52 N1911A/1912A P-Series Power Meters Programming Guide
Editing Frequency Dependent Offset Tables
It is not possible to create any additional frequency dependent offset tables.
However, the 10 existing tables can be edited using the MEMory subsystem.
To do this:
1 Select one of the existing tables using:
MEMory:TABle:SELect <string>
For information on naming frequency dependent offset tables see “Naming
Frequency Dependent Offset Tables” on page 54. For information on the
current names which you can select refer to “Listing the Frequency Dependent
Offset Table Names” on page 53.
2 Enter the frequency data using:
MEMory:TABle:FREQuency <numeric_value> {,<numeric_value>}
3 Enter the offset factors as shown in the table below using:
MEMory:TABle:GAIN <numeric_value> {,<numeric_value>}
4 If required, rename the frequency dependent offset table using:
MEMory:TABLe:MOVE <string>,<string>. The first <string> parameter
identifies the existing table name, and the second identifies the new table
name.
Frequency Offset
Frequency 1 Offset 1
Frequency 2 Offset 2
""
Frequency n Offset n