EasyManua.ls Logo

Keysight N1911A

Keysight N1911A
771 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 319
PSTatistic:CCDF:GAUSsian[:STATe] <boolean>
This command is used to turn on or off the Gaussian trace and it is independent of
the channels attached.
Syntax
Example
Reset Conditions
On reset, the Gaussian trace will be cleared (OFF).
NOTE
This command is only applicable when P-Series sensors are present and the
following conditions are met:
Free-run acquisition mode is selected
NORMal or DOUBle measurement speed setting is chosen
Expanded window is enabled (not full CCDF screen window)
0|OFF
1|ON
Space
?
:STAT
:GAUS
:CCDF
PST
PST:CCDF:GAUS ON This command turns on the Gaussian trace.
PST:CCDF:GAUS OFF This command turns off the Gaussian trace.

Table of Contents

Other manuals for Keysight N1911A

Related product manuals