1 Power Meter Remote Operation
66 N1911A/1912A P-Series Power Meters Programming Guide
Using STATus
If using GPIB, you can use the STATus subsystem to generate an SRQ to interrupt
your program when a limit failure occurs. This is a more efficient method than
using SENSe or CALCulate, since you do not need to check the limit failures after
every power measurement.
Refer to “Status Reporting” on page 73 and “STATus Subsystem” on page 465 for
further information.
If TRIGger:DELay:AUTO is set to ON, then the number of failures returned by
SENSe:LIMit:FCOunt? or CALCulate[1|2]:LIMit:FCOunt? is affected by the
current filter settings.