SERVice Subsystem 15
N1911A/1912A P-Series Power Meters Programming Guide 701
SERVice:OPTion <character_data>
This command loads the power meter memory with the options fitted. The query
form of the command can be used to determine which options are fitted to the
unit.
Syntax
Parameters
Example
Query
SERVice:OPTion?
The query returns the current option string. For example, if the string “003” is
returned, the power meter is fitted with a sensor input and power reference on the
back panel.
Item Description/Default Range of Values
character_data Details the option number in a comma separated list.
A maximum of 30 characters can be used.
A to Z (uppercase)
a to z (lowercase)
0 - 9
_ (underscore)
SERV:OPT “003” This command loads the power meter
memory with 003 indicating that the unit is a
rear panel option.
SERV
:OPT
character_data
?
Space

Table of Contents