Loading...
OUTPut Subsystem 8
N1911A/1912A P-Series Power Meters Programming Guide 311
OUTPut:RECorder[1]|2:STATe <boolean>
This command enables or disables the specified recorder output.
Syntax
Example
Reset Condition
On reset, the recorder output is OFF.
Query
OUTPut:RECorder[1]|2:STATe?
The query command enters a 1 or 0 into the output buffer indicating whether or
not the specified recorder is switched on.
1 is returned when the recorder output is switched ON
0 is returned when the recorder output is switched OFF
Query Example
OUTP:REC1:STAT 1 This command enables the specified
recorder output.
OUTP
:STAT
:REC
0|OFF
1|ON
Space
?
1
2
OUTP:REC2:STAT? This command queries the status of the
recorder output.