FORMat Subsystem 6
N1911A/1912A P-Series Power Meters Programming Guide 265
FORMat[:READings][:DATA] <character_data>
This command sets the data format for transferring numeric information to either
ASCii or REAL:
– When the format type is ASCii, numeric data is output as ASCII bytes in the
<NR3> format.
– When the format type is REAL, numeric data is output as IEEE 754 64 bit
floating point numbers in a definite length block. The result is an 8 byte block
per number. Each complete block is terminated by a line feed character.
For the N1912A power meter the same FORMat is used on both channels.
Syntax
Parameters
NOTE
FORMat data formatting is not affected by TRACe subsystem data formatting.
Item Description/Default Range of Values
character_data Data format for transferring data:
– ASCii
–REAL
ASCii
REAL
FORM
:READ :DATA
Space
?
character_data