7 MEMory Subsystem
298 N1911A/1912A P-Series Power Meters Programming Guide
MEMory:TABLe:GAIN[:MAGNitude]:POINts?
This query is used to return the number of calibration factor/offset points for the
currently selected table. If the currently selected table is a sensor calibration table
(8480 Series sensors and N8480 Series sensors with Option CFT only), the
reference calibration factor is included
If no values have been set, 0 is returned. If no table is selected, NAN is returned.
Syntax
Example
MEM:TABL:GAIN:POIN? This command queries the number of
calibration factor/offset points in the current
table.