PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 351
PSTatistic[1]|2:CCDF:STORe:REFerence
This command is used to store Channel A or Channel B as a reference trace for
CCDF graph window.
Syntax
Example
NOTE
The trace will be saved as reference trace in volatile RAM.
NOTE
This command is only applicable when P-Series sensors are present and the
following conditions are met:
Free-run acquisition mode is selected
NORMal or DOUBle measurement speed setting is chosen
Expanded window is enabled (not full CCDF screen window)
PST
2
1
:CCDF
:STOR
:REF
PST:CCDF:STOR:REF This command saves the Channel A trace as
reference trace.
PST2:CCDF:STOR:REF This command saves the Channel B trace as
reference trace.