9 PSTatistic Subsystem
352 N1911A/1912A P-Series Power Meters Programming Guide
Reset Condition
On reset, the previously saved reference trace will be cleared.
Query
PSTatistic[1]|2:CCDF:STORe:REFerence?
The query enters a 1 or 0 into the output buffer indicating the status of the CCDF
reference.
1 is returned when there is a saved reference trace
0 is returned when there is no saved reference trace
Query Eample
Error Messages
If no power sensor is connected, error –241 "Hardware missing" occurs.
If sensor/sensors connected are not P-Series sensors, error –241 "Hardware
missing" occurs.
If measurement speed setting is FAST, error –221 "Settings conflict" occurs.
If command is executed in other window besides expanded CCDF window,
error –221 "Settings conflict: Requires CCDF expanded window" occurs.
If the acquisition mode is in continuous triggering or triggering source is set to
either INT1, INT2, or EXT, error –221 "Settings conflict" occurs.
PST:CCDF:STOR:REF? This command queries whether there is
saved reference trace or not.