15 SERVice Subsystem
708 N1911A/1912A P-Series Power Meters Programming Guide
SERVice:SENSor[1]|2:CORRections:STATe <boolean>
This command enables/disables the voltage to corrected power conversion. It
applies to E9320 Series power sensors only.
Syntax
Example
Reset Condition
On reset, the value is set to ON.
NOTE
Before setting this command to OFF, you must set the INIT:CONT command to
OFF.
After setting this command to OFF, you must only run commands relating to the
gathering of ADC values—for example, the SERV:SENS:RADC command.
SERV:SENS2:CORR:STAT ON This command enables the voltage to
corrected power conversion for Channel B.
0|OFF
1|ON
Space
?
:STAT
:CORR
SERV
1
2
:SENS