2 MEASurement Commands
112 N1911A/1912A P-Series Power Meters Programming Guide
Figure 2-1 Measurement Display CALCulate Block Window
Optional Parameters
CONFigure, FETCh?, READ? and MEASure? have the following three optional
parameters:
– An expected power value
–A resolution
– A source list
Expected Power Value
An <expected_value> parameter is only required if you are using an E-Series
power sensor or N8480 Series power sensor (excluding Option CFT). It has no
effect on P-Series power sensor, 8480 Series power sensor or N8480 Series
power sensor with Option CFT. The value entered determines which of the power
sensor’s two ranges is used for the measurement. If the current setting of the
power sensor’s range is no longer valid for the new measurement, specifying the
expected power value decreases the time taken to obtain a result.
upper window/upper measurement
CONfigure1?
upper window/lower measurement
CONFigure3?
lower window/upper measurement
CONFigure2?
lower window/lower measurement
CONFigure4?