SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 417
Parameters
[a] The following measurement units can be used:
– Hz
– kHz (10
3
)
– MHz (10
6
)
– GHz (10
9
)
Example
Reset Condition
On reset, the frequency is set to 50 MHz (DEF).
Query
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed]? [MIN|MAX]
The query returns the current frequency setting or the values associated with MIN
and MAX. The units in which the results are returned are Hz.
Query Example
Item Description/Default Range of Values
numeric_value A numeric value for the frequency:
–
DEF: the default value is 50 MHz
– MIN: 1 kHz
–
MAX: 1000.0 GHz
The default units are Hz.
1 kHz to 1000.0 Ghz
[a]
DEF
MIN
MAX
FREQ 500kHz This command enters a Channel A frequency
of 500 kHz.
SENS2:FREQ? This command queries the Channel B
frequency setting.